Nitrous oxide (N2O) gas detector
Fixed nitrous oxide (N2O) detectors for operating theatres, dental surgeries, medical gas plants and the food industry: NDIR infrared sensor, 0-1,000 ppm, lifetime over 5 years. Choose the model below to see its specifications and options.
Fixed nitrous oxide (N2O) detectors for operating theatres, dental surgeries, medical gas plants and the food industry: NDIR infrared sensor, 0-1,000 ppm, lifetime over 5 years.
Why detect N2O?
Nitrous oxide is an anaesthetic gas used in operating theatres, maternity wards and dental surgeries, and a propellant in the food industry. Prolonged staff exposure must be avoided: detection monitors the rooms and the cylinder stores.
The sensor
NDIR infrared sensor, 0-1,000 ppm range, 1 ppm resolution, operation from −20 to +50 °C, lifetime over 5 years, checked every 12 months.
Where to install an N2O detector?
At breathing height in the rooms, and close to the floor in storage rooms, as N2O is heavier than air.
Available models
| Model | Sensor | Ranges | Housing | Outputs | Local alarm |
|---|---|---|---|---|---|
| E2638-N2O | NDIR infrared | 0-1,000 ppm | grey ABS, IP65, 90 × 145 × 55 mm | 2 × 4-20 mA / 0-10 V, RS485 Modbus RTU | optional: 2 relays 5 A, LEDs and 85 dB buzzer or LCD display |
| E2639-N2O | NDIR infrared | 0-1,000 ppm | grey ABS, IP65, 190 mm wide | 2 × 4-20 mA, RS485 Modbus RTU, optional WiFi | 3 relays (pre-alarm, alarm, fault), 4 LEDs, 85 dB buzzer, 2 × 24 V beacon outputs |
| E2608-N2O | NDIR infrared | 0-1,000 ppm | grey ABS, IP65, 87 × 82 × 55 mm, wall or duct | 2 × 4-20 mA / 0-10 V, RS485 Modbus RTU | 2 SPST relays 5 A |
| E2618-N2O | NDIR infrared | 0-1,000 ppm | grey ABS, IP65, 82 × 80 × 55 mm, wall or duct | 2 × 4-20 mA / 0-10 V, RS485 Modbus RTU | none: alarm handled by the PLC |
| E2648-N2O | NDIR infrared | 0-1,000 ppm | die-cast aluminium, IP66, 120 × 125 × 57 mm | 2 × 4-20 mA / 0-10 V, RS485 Modbus RTU | optional 2 SPST relays 5 A |
Choose the model in the list at the top of the page to see all its specifications and options. Other measuring ranges are available on request.
Alarm and options by model
- E2638 : optional: 2 relays 5 A, LEDs and 85 dB buzzer or LCD display ; relays, buzzer and LEDs, built-in or remote LCD display (0 to +50 °C), self-test button, 3 m remote probe, 90-265 VAC.
- E2639 : 3 relays (pre-alarm, alarm, fault), 4 LEDs, 85 dB buzzer, 2 × 24 V beacon outputs ; WiFi with web server, replaceable pre-calibrated GSH sensor, 2.5 or 5 m remote probe (GSH version).
- E2608 : 2 SPST relays 5 A ; duct version, 3 m remote probe, 24 VAC or 230 VAC.
- E2618 : none: alarm handled by the PLC ; duct version, 3 m remote probe.
- E2648 : optional 2 SPST relays 5 A ; relays, 3 m remote probe, 24 VAC or 230 VAC.
Request a quote
Afric Froid supplies and installs Evikon gas detectors with its refrigeration systems in Tunisia, Algeria, Senegal and Côte d’Ivoire. Tell us the room, its temperature, the number of points to monitor, the power supply and how the alarm should be given: we select the model and send you a quote.
See all our gas detectors and sensors.
| Model | E2638-N2O · multi-option IP65, E2639-N2O · 3 relays, buzzer, E2608-N2O · 2 relays, wall/duct, E2618-N2O · transmitter, no relays, E2648-N2O · aluminium IP66 |
|---|
Official Evikon documents, kept up to date by the manufacturer:
Related products
Combustible gas detector (methane, LPG, propane)
Fixed combustible and explosive gas detectors for boiler rooms, industrial kitchens, propane rooms and machinery rooms: methane, natural gas, propane, butane, LPG, hydrogen, measured from 0 to 100 % LEL. Stand-alone alarms or 4-20 mA transmitters. Choose the model below to see its specifications and options.